Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sanwa CD-780C: (1 year) CAL VER /5520 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 30 NUMBER OF LINES: 127 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Sanwa Service Manual # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 30 min. - REQUIRED 1.005 DISP Temperature: 22 - 24 degC 1.005 DISP Humidity: <70% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {INITIAL SETUP} 1.008 DISP Turn the UUT on. 1.009 DISP Connect the 5520A to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.009 DISP [32] 5520A NORMAL LO to UUT COM 1.010 HEAD {DC VOLTAGE} 1.011 HEAD DC VOLTAGE: {200mV Range} 1.012 5520 200 0.0mV 0.2U 2W 2.001 5520 200 190.0mV 1.2U 2W 3.001 HEAD DC VOLTAGE: {2V Range} 3.002 5520 2 1.900V 0.02U 2W 4.001 5520 2 -1.900V 0.02U 2W 5.001 HEAD DC VOLTAGE: {20V Range} 5.002 5520 20 19.00V 0.17U 2W 6.001 HEAD DC VOLTAGE: {200V Range} 6.002 5520 200 190.0V 1.7U 2W 7.001 HEAD DC VOLTAGE: {1000V Range} 7.002 5520 1000 1000V 9U 2W 8.001 HEAD {AC VOLTAGE} 8.002 HEAD AC VOLTAGE: {2V Range} 8.003 5520 2 1.900V 0.034U 60H SI 2W 9.001 HEAD AC VOLTAGE: {20V Range} 9.002 5520 20 19.00V 0.34U 60H SI 2W 10.001 HEAD AC VOLTAGE: {200V Range} 10.002 5520 200 190.0V 3.4U 60H SI 2W 11.001 HEAD AC VOLTAGE: {750V Range} 11.002 5520 750 700V 11U 60H SI 2W 12.001 HEAD {RESISTANCE} 12.002 HEAD RESISTANCE: {20 MOhm Range} 12.003 5520 20 10MZ 0.62U 2W 13.001 HEAD RESISTANCE: {2000 kOhm Range} 13.002 5520 2000 1900kZ 32U 2W 14.001 HEAD RESISTANCE: {200 kOhm Range} 14.002 5520 200 190.0kZ 1.7U 2W 15.001 HEAD RESISTANCE: {20 kOhm Range} 15.002 DISP Connect the 5520A to the UUT as follows: 15.002 DISP [32] 5520A AUX HI to UUT V/OHM 15.002 DISP [32] 5520A AUX LO to UUT COM 15.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 15.002 DISP [32] 5520A NORMAL LO to UUT COM 15.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 15.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 15.003 5520 20 19.00kZ 0.17U CW 16.001 HEAD RESISTANCE: {2 kOhm Range} 16.002 5520 2 1.900kZ 0.017U CW 17.001 HEAD RESISTANCE: {200 Ohm Range} 17.002 5520 200 190Z 1.7U CW 18.001 DISP Select CONTINUITY. 18.002 5520 0.0Z S CW 18.003 EVAL {Is the buzzer ON (0 ohm)}? 19.001 5520 2.1kZ S CW 19.002 EVAL {IS the buzzer OFF (>2k ohm)}? 20.001 HEAD {DC LOW CURRENT} 20.002 DISP Connect the 5520A to the UUT as follows: 20.002 DISP [32] 5520A AUX HI to UUT uA/mA 20.002 DISP [32] 5520A AUX LO to UUT COM 20.003 HEAD DC LOW CURRENT: {200uA Range} 20.004 5520 200 190.0uA 0.2U 2W 21.001 HEAD DC LOW CURRENT: {2mA Range} 21.002 5520 2 1.900mA 0.002U 2W 22.001 HEAD DC LOW CURRENT: {20mA Range} 22.002 5520 20 19.00mA 0.02U 2W 23.001 HEAD DC LOW CURRENT: {200mA Range} 23.002 5520 200 190.0mA 0.2U 2W 24.001 HEAD {AC LOW CURRENT} 24.002 HEAD AC LOW CURRENT: {200uA Range} 24.003 5520 200 190.0uA 0.4U 60H SI 2W 25.001 HEAD AC LOW CURRENT: {2mA Range} 25.002 5520 2 1.900mA 0.004U 60H SI 2W 26.001 HEAD AC LOW CURRENT: {20mA Range} 26.002 5520 20 19.00mA 0.04U 60H SI 2W 27.001 HEAD AC LOW CURRENT: {200mA Range} 27.002 5520 200 190.0mA 0.4U 60H SI 2W 28.001 HEAD {AC HIGH CURRENT} 28.002 DISP Connect the 5520A to the UUT as follows: 28.002 DISP [32] 5520A 20A to UUT 20A 28.002 DISP [32] 5520A AUX LO to UUT COM 28.003 HEAD AC HIGH CURRENT: {20A Range} 28.004 5520 20 10.00A 0.169U 60H SI HC 2W 29.001 HEAD {DC HIGH CURRENT} 29.002 HEAD DC HIGH CURRENT: {20A Range} 29.003 5520 20 10.00A 0.210U HC 2W 30.001 END